Autoclaving plastics: what to watch for in steam sterilization

Not every plastic survives the autoclave. This guide shows, in a table taken from Semadeni's chemical resistance list, which of 20 plastics can be autoclaved or sterilized with dry heat, ethylene oxide, gamma rays or chemicals, and up to which temperature each can be used continuously. It also covers the recommended cycle, why caps must come off and how to spot fatigued parts.

Autoclaving plastics: what to watch for in steam sterilization

Which plastics can be autoclaved?

According to Semadeni's resistance list, PP, PC, PMP, PSU, PES, PVDF, PTFE, PFA, FEP, ETFE, ECTFE, POM and silicone can be autoclaved at 121 °C, 1 bar for 20 minutes. PE-HD, PE-LD, PS, PMMA, PET, PVC, ABS, SAN and PA cannot. PC and PSU lose strength after repeated autoclaving. Caps must be fully opened during autoclaving.

recommended autoclave cycle
121 °C · 1 bar · 20 min
otherwise the vessel deforms inwards
Caps off
lose strength after repeated autoclaving
PC, PSU
not autoclavable, continuous use up to +90 °C
PE-HD

Open the resistance list (PDF in German)

What happens in the autoclave

Autoclaving is sterilization with saturated steam under pressure. Semadeni's resistance list recommends a cycle of 20 minutes at 121 °C and 1 bar for plastic labware. Some applications also run at 134 °C.

Moist heat kills microorganisms and stresses the plastic at the same time. Whether a material copes depends mainly on the temperature up to which it stays dimensionally stable. PE-HD, for example, can be used continuously up to +90 °C and briefly up to +110 °C, so it is not autoclavable. PP withstands +120 °C continuously and +140 °C briefly.

The table below states what the material tolerates. Whether a specific item may be autoclaved also depends on wall thickness, design and closure. The information given for the item is decisive.

What to watch for when autoclaving

  1. Open the closure: Always unscrew caps before autoclaving and place them loosely on the container. A closed vessel deforms inwards when the pressure changes, and this cannot be reversed. Only screw the cap on tightly once the item has cooled.
  2. Clean first: Clean items thoroughly and rinse them with distilled water before autoclaving. Chemical residues that are harmless at room temperature can attack the plastic at 121 °C. Do not autoclave containers holding detergent (except fluoropolymer containers).
  3. Allow longer heat-up times: Plastic conducts heat less well than glass or metal. Liquids in large plastic containers take longer to reach 121 °C. The required sterilization time can only be determined by your own measurements.
  4. Check parts after each cycle: Strength can decrease after repeated autoclaving, especially with PC and PSU. Check for stress cracks and defects before reuse. Frequent steam sterilization can make PC centrifuge tubes unusable.

Which plastic tolerates which sterilization method?

Open the resistance list (PDF in German)

Plastic
Steam
Dry heat
EtO
Gamma
Continuous max.
PP
✓
✗
✓
✗
+120 °C
PE-HD
✗
✗
✓
✓
+90 °C
PE-LD
✗
✗
✓
✓
+80 °C
PC
✓²
✗
✓
✗
+135 °C
PMP (TPX)
✓
✓³
✓
✗
+130 °C
PSU
✓
✓
✓
✓
+160 °C
PES
✓
✓
✓
✓
+200 °C
PET, copolyester
✗
✗
✓
✓
+60 °C
PS
✗
✗
✓
✓
+70 °C
PMMA
✗
✗
✓
✓
+85 °C
PVC rigid
✗
✗
✓
✗
+70 °C
PA
✗
✗
✓
✓
+100 °C
POM
✓³
✗
✓
✗
+110 °C
PVDF
✓
✗
✓
✗
+105 °C
ECTFE
✓
✓¹
✓
✗
+150 °C
ETFE
✓
✓¹
✓
✗
+150 °C
FEP
✓
✓
✓
✗
+205 °C
PFA
✓
✓
✓
✗
+260 °C
PTFE
✓
✓
✓
✗
+260 °C
Silicone (SI)
✓
✓
✓
✗
+170 °C

✓ = suitable · ✗ = not suitable · Steam = autoclave at 121 °C, 1 bar, 20 min · Dry heat = 160 °C · EtO = gas sterilization with ethylene oxide · Continuous max. = continuous service temperature of the material. ¹ only up to 150 °C · ² frequent steam sterilization leads to loss of strength, PC centrifuge tubes can become unusable · ³ sterilization reduces mechanical strength. According to the list, all plastics shown can be sterilized chemically (formalin, ethanol); aggressive agents can cause hairline cracks with prolonged exposure. Source: Semadeni resistance list, tables «Sterilisation von Kunststoffen» and «Verhalten bei Temperatureinwirkung». The data refer to the material; for an individual item, its stated service temperature applies.

Practical recommendations

  • If labware is autoclaved regularly: PP is the usual material: autoclavable and, according to the list, usable continuously up to +120 °C. For more temperature headroom, consider PMP, PSU or fluoropolymers (PFA, FEP, PTFE).
  • If the part must be crystal clear: PC and PMP (TPX) are transparent and autoclavable. PC loses strength with frequent steam sterilization. PMMA and PS are not autoclavable.
  • If only PE-HD is an option: PE-HD is not autoclavable. According to the list it can be sterilized with ethylene oxide, chemically or with gamma rays.
  • If the part is cloudy after autoclaving: Some transparent plastics absorb a little water vapour and are cloudy afterwards. The haze disappears as the plastic dries, faster in a drying oven at +110 °C. If it remains, chemical additives in the steam may be the cause.

Limits of the table

The table rates materials, not items. The range of use of a finished part also depends on design, wall thickness, operating conditions and contents; the service temperature stated for the item is decisive. The list does not say how often an item can be autoclaved. Autoclavable does not mean chemically resistant: media that are harmless at room temperature can attack at 121 °C. The information is a recommendation without guarantee.

Autoclaving: moist heat under pressure

In the autoclave, saturated steam acts on the items being sterilized. For plastic labware, Semadeni’s resistance list recommends 20 minutes at 121 °C and 1 bar. The rule of thumb from the table above: PP, PC, PMP, PSU and the fluoropolymers are autoclavable; polyethylene, polystyrene, acrylic and PET are not.

What matters is the temperature a material withstands continuously. If it is well below 121 °C, the part deforms. That rules out PE-HD (continuous +90 °C), PE-LD (+80 °C), PS (+70 °C) and PMMA (+85 °C). PVDF (+105 °C) and POM (+110 °C) are still autoclavable because they briefly tolerate considerably more (+150 °C).

Prepare, load, cool down

  1. Clean and rinse. Residues of chemicals and detergent attack the plastic at 121 °C or bake on. Rinse with distilled water.
  2. Unscrew caps. Place the cap loosely on the vessel. This applies to every sterilization method except gamma rays.
  3. Stand racks level. Autoclave test-tube racks on a flat surface, otherwise they warp.
  4. Let it cool. Only screw caps on tightly once cooled.
  5. Inspect. Before reuse, check for stress cracks, warping and haze, especially with PC and PSU.

Other sterilization methods

Ethylene oxide (EtO): All plastics in the table can be gas-sterilized. Depending on the application, let items air out sufficiently afterwards. Here too, unscrew caps completely, as pressure differences arise at higher temperatures.

Dry heat (160 °C): Only for materials with high temperature headroom: PSU, PES, PMP, silicone and fluoropolymers. PP and PC are autoclavable but not dry-heat resistant.

Chemical: Most plastics tolerate common disinfectants such as quaternary ammonium compounds, iodine compounds, formalin or ethanol. Aggressive agents can cause hairline cracks with prolonged exposure, especially on PS, SAN, PVC, PC and PMMA. Which agent attacks which plastic is shown in the chemical resistance chart.

Gamma rays: Sterilization takes place at room temperature via a high radiation dose. According to the list, suitable materials include PE-HD, PE-LD, PS, PET and PSU, but not PP, PC or the fluoropolymers.

Microwave: In general, plastics are transparent to microwaves, PMP and PSU the most. For aggressive chemicals such as acids and solvents, use fluoropolymer containers and ensure extraction. Remove caps beforehand.

Choosing the right material

For bottles, beakers and containers that go into the autoclave regularly, PP is the usual material. Where the vessel must be crystal clear, PC and PMP are the alternatives; PC loses strength with frequent autoclaving. For higher temperatures, dry heat or aggressive media, PSU and the fluoropolymers PFA, FEP and PTFE are options.

Whether a specific item is autoclavable also depends on its design and wall thickness. The service temperature stated for the item is decisive. Items for sterilization in the lab can be found in the shop under Sterilization, autoclaving, disinfection.

Sources

Frequently asked questions

Question not listed?

We're happy to answer it personally.

Get in touch
At what temperature are plastics autoclaved?

Semadeni's resistance list recommends 20 minutes at 121 °C and 1 bar. Liquids in large plastic containers take longer to reach this temperature because plastic conducts heat less well than glass. The required time can only be determined by your own measurements.

Which plastics are autoclavable?

According to the list: PP, PC, PMP (TPX), PSU, PES, PVDF, POM, silicone and the fluoropolymers PTFE, PFA, FEP, ETFE and ECTFE. With PC and POM, sterilization reduces strength. Not autoclavable: PE-HD, PE-LD, PS, PMMA, PET and copolyester, PVC, ABS, SAN and PA.

Can HDPE be autoclaved?

No. According to the resistance list, PE-HD can be used continuously up to +90 °C and briefly up to +110 °C, so it is not suitable for 121 °C. It can be sterilized with ethylene oxide, chemically or with gamma rays.

Can polycarbonate be autoclaved?

Yes, but not indefinitely. Frequent steam sterilization causes PC to lose strength; PC centrifuge tubes can become unusable. Check PC parts for stress cracks before each reuse.

Why do bottles deform in the autoclave?

Usually because the cap was closed. When the pressure changes, a closed vessel is pulled inwards, and this deformation remains. Unscrew caps, place them loosely on the vessel and only close them after cooling.

Is autoclavable the same as dry-heat sterilizable?

No. According to the list, only PSU, PES, PMP, silicone and the fluoropolymers tolerate dry heat at 160 °C, ECTFE and ETFE only up to 150 °C. PP and PC are autoclavable but not suitable for dry heat.

Which plastics can be sterilized with gamma rays?

According to the list, among others PE-HD, PE-LD, PS, PMMA, PET, PA, PSU and PES. PP, PC, POM, PVC, PVDF, silicone and the fluoropolymers are not suitable for gamma sterilization.

Related

Our services

From development and injection moulding to assembly – technical and food-safe solutions from our own Swiss production.

Explore our services

A project or a question?

Talk to us

We assess feasibility, advise on material and standards and prepare a matching quote for your project.

Get in touch

Never miss an update

Expert articles on regulation and materials, product news and insights – occasional and spam-free.

By clicking “Subscribe”, you sign up for the Semadeni AG e-mail newsletter, published up to six times a year. I agree that the above information together with my IP address, a date stamp and any other metadata may be transmitted to Semadeni AG, where it will be stored and further processed for the purpose of processing my request. You can unsubscribe from these communications at any time. For more information on how to unsubscribe, our privacy practices, and how we are committed to protecting and respecting your privacy, please review our privacy policy. By clicking “Subscribe” below, you consent to allow Semadeni AG to store and process the personal information submitted above to provide you the content requested.

Which item will survive your sterilization cycle?

Tell us the method, temperature, holding time, how often you sterilize and which media are involved. We will tell you what the resistance list and our suppliers' data say about it and which items from our range are suitable. Our information is a recommendation without guarantee.

Prefer personal advice