Autoclaving plastics: what to watch for in steam sterilization
Not every plastic survives the autoclave. This guide shows, in a table taken from Semadeni's chemical resistance list, which of 20 plastics can be autoclaved or sterilized with dry heat, ethylene oxide, gamma rays or chemicals, and up to which temperature each can be used continuously. It also covers the recommended cycle, why caps must come off and how to spot fatigued parts.
Which plastics can be autoclaved?
According to Semadeni's resistance list, PP, PC, PMP, PSU, PES, PVDF, PTFE, PFA, FEP, ETFE, ECTFE, POM and silicone can be autoclaved at 121 °C, 1 bar for 20 minutes. PE-HD, PE-LD, PS, PMMA, PET, PVC, ABS, SAN and PA cannot. PC and PSU lose strength after repeated autoclaving. Caps must be fully opened during autoclaving.
- recommended autoclave cycle
- 121 °C · 1 bar · 20 min
- otherwise the vessel deforms inwards
- Caps off
- lose strength after repeated autoclaving
- PC, PSU
- not autoclavable, continuous use up to +90 °C
- PE-HD
Open the resistance list (PDF in German)
What happens in the autoclave
Autoclaving is sterilization with saturated steam under pressure. Semadeni's resistance list recommends a cycle of 20 minutes at 121 °C and 1 bar for plastic labware. Some applications also run at 134 °C.
Moist heat kills microorganisms and stresses the plastic at the same time. Whether a material copes depends mainly on the temperature up to which it stays dimensionally stable. PE-HD, for example, can be used continuously up to +90 °C and briefly up to +110 °C, so it is not autoclavable. PP withstands +120 °C continuously and +140 °C briefly.
The table below states what the material tolerates. Whether a specific item may be autoclaved also depends on wall thickness, design and closure. The information given for the item is decisive.
What to watch for when autoclaving
- Open the closure: Always unscrew caps before autoclaving and place them loosely on the container. A closed vessel deforms inwards when the pressure changes, and this cannot be reversed. Only screw the cap on tightly once the item has cooled.
- Clean first: Clean items thoroughly and rinse them with distilled water before autoclaving. Chemical residues that are harmless at room temperature can attack the plastic at 121 °C. Do not autoclave containers holding detergent (except fluoropolymer containers).
- Allow longer heat-up times: Plastic conducts heat less well than glass or metal. Liquids in large plastic containers take longer to reach 121 °C. The required sterilization time can only be determined by your own measurements.
- Check parts after each cycle: Strength can decrease after repeated autoclaving, especially with PC and PSU. Check for stress cracks and defects before reuse. Frequent steam sterilization can make PC centrifuge tubes unusable.
Which plastic tolerates which sterilization method?
Open the resistance list (PDF in German)
✓ = suitable · ✗ = not suitable · Steam = autoclave at 121 °C, 1 bar, 20 min · Dry heat = 160 °C · EtO = gas sterilization with ethylene oxide · Continuous max. = continuous service temperature of the material. ¹ only up to 150 °C · ² frequent steam sterilization leads to loss of strength, PC centrifuge tubes can become unusable · ³ sterilization reduces mechanical strength. According to the list, all plastics shown can be sterilized chemically (formalin, ethanol); aggressive agents can cause hairline cracks with prolonged exposure. Source: Semadeni resistance list, tables «Sterilisation von Kunststoffen» and «Verhalten bei Temperatureinwirkung». The data refer to the material; for an individual item, its stated service temperature applies.
Practical recommendations
- If labware is autoclaved regularly: PP is the usual material: autoclavable and, according to the list, usable continuously up to +120 °C. For more temperature headroom, consider PMP, PSU or fluoropolymers (PFA, FEP, PTFE).
- If the part must be crystal clear: PC and PMP (TPX) are transparent and autoclavable. PC loses strength with frequent steam sterilization. PMMA and PS are not autoclavable.
- If only PE-HD is an option: PE-HD is not autoclavable. According to the list it can be sterilized with ethylene oxide, chemically or with gamma rays.
- If the part is cloudy after autoclaving: Some transparent plastics absorb a little water vapour and are cloudy afterwards. The haze disappears as the plastic dries, faster in a drying oven at +110 °C. If it remains, chemical additives in the steam may be the cause.
Limits of the table
The table rates materials, not items. The range of use of a finished part also depends on design, wall thickness, operating conditions and contents; the service temperature stated for the item is decisive. The list does not say how often an item can be autoclaved. Autoclavable does not mean chemically resistant: media that are harmless at room temperature can attack at 121 °C. The information is a recommendation without guarantee.
Autoclaving: moist heat under pressure
In the autoclave, saturated steam acts on the items being sterilized. For plastic labware, Semadeni’s resistance list recommends 20 minutes at 121 °C and 1 bar. The rule of thumb from the table above: PP, PC, PMP, PSU and the fluoropolymers are autoclavable; polyethylene, polystyrene, acrylic and PET are not.
What matters is the temperature a material withstands continuously. If it is well below 121 °C, the part deforms. That rules out PE-HD (continuous +90 °C), PE-LD (+80 °C), PS (+70 °C) and PMMA (+85 °C). PVDF (+105 °C) and POM (+110 °C) are still autoclavable because they briefly tolerate considerably more (+150 °C).
Prepare, load, cool down
- Clean and rinse. Residues of chemicals and detergent attack the plastic at 121 °C or bake on. Rinse with distilled water.
- Unscrew caps. Place the cap loosely on the vessel. This applies to every sterilization method except gamma rays.
- Stand racks level. Autoclave test-tube racks on a flat surface, otherwise they warp.
- Let it cool. Only screw caps on tightly once cooled.
- Inspect. Before reuse, check for stress cracks, warping and haze, especially with PC and PSU.
Other sterilization methods
Ethylene oxide (EtO): All plastics in the table can be gas-sterilized. Depending on the application, let items air out sufficiently afterwards. Here too, unscrew caps completely, as pressure differences arise at higher temperatures.
Dry heat (160 °C): Only for materials with high temperature headroom: PSU, PES, PMP, silicone and fluoropolymers. PP and PC are autoclavable but not dry-heat resistant.
Chemical: Most plastics tolerate common disinfectants such as quaternary ammonium compounds, iodine compounds, formalin or ethanol. Aggressive agents can cause hairline cracks with prolonged exposure, especially on PS, SAN, PVC, PC and PMMA. Which agent attacks which plastic is shown in the chemical resistance chart.
Gamma rays: Sterilization takes place at room temperature via a high radiation dose. According to the list, suitable materials include PE-HD, PE-LD, PS, PET and PSU, but not PP, PC or the fluoropolymers.
Microwave: In general, plastics are transparent to microwaves, PMP and PSU the most. For aggressive chemicals such as acids and solvents, use fluoropolymer containers and ensure extraction. Remove caps beforehand.
Choosing the right material
For bottles, beakers and containers that go into the autoclave regularly, PP is the usual material. Where the vessel must be crystal clear, PC and PMP are the alternatives; PC loses strength with frequent autoclaving. For higher temperatures, dry heat or aggressive media, PSU and the fluoropolymers PFA, FEP and PTFE are options.
Whether a specific item is autoclavable also depends on its design and wall thickness. The service temperature stated for the item is decisive. Items for sterilization in the lab can be found in the shop under Sterilization, autoclaving, disinfection.